ABL800 FLEX Operator's Manual
9. Data management
Calibration Log
Purpose
The Calibration Log is a historical file of calibrations automatically saved in the
Calibration Log after a calibration.
Each calibration report is listed as an individual record. The reports are listed
chronologically, the latest report being at the top of the screen.
Each report is identified as follows (use the left/right scroll bar to see the items):
Item Description
Time
Date and time the calibration was performed.
Cal #
The ID number of a performed calibration.
Type
Indicates the calibration type: 1-point, 2-point or tHb.
Status
Indicates the calibration status:
OK
a successful calibration
?
an error detected or a parameter exceeded the
measuring range
Interrupted
a calibration stopped by the operator
Aborted
a calibration stopped by the analyzer most
likely due to missing calibration solution.
NOTICE:
The columns displayed in the log and their positions can be changed.
Contact a qualified service technician.
Continued on next page
9-17
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...