ABL800 FLEX Operator's Manual
9. Data management
Quality Control Log,
Continued
To view the
Quality Control Plot
screen press
Plot
on the
Quality Control Log
screen.
Quality control
plot
The following diagram shows the details of a typical plot.
Time
Measurement
Slot number
Solution type
Out-of-range symbol
Control range upper limit
Control range lower limit
Measured value
Date
Lot change
Parameter
Plot element
Function
Shaded block
Date, time and measured value for the highlighted
measurement. Use the arrow buttons to scroll the plot and
view other measurements.
Control range
limits
Show the upper and the lower limits of the control range
for the highlighted measurement.
Dots
Show the number of measurements for the selected
parameter. To view a measurement result, highlight a dot
on the screen and press
Result
.
Out-of-range
symbols
Measurement outside the control range and
inside the statistical range.
Measurement outside both the control and
statistical ranges.
Continued on next page
9-15
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...