15. Radiometer settings
ABL800 FLEX Operator's Manual
15-6
Radiometer default settings,
Continued
Temperature Corrected Values
pH(
T
)
p
CO
2
(
T
)
p
O
2
(
T
)
< New Group >
Oxygen Status
c
tO
2
p
50
< New Line >
Acid Base Status
c
Base(Ecf)
)
st
P,
(
3
HCO
c
Default settings for
Drift tolerances
are as follows:
Parameter
Drift 1
Drift 2
Units
pH 0.020
0.020
p
CO
2
2.5 5 mmHg
p
O
2
6 6 mmHg
c
tHb 10 N/A pA
c
K
+
0.2 1.5
mmol/L
c
Na
+
3 1 mmol/L
c
Ca
2+
0.05 0.2 mmol/L
c
Cl
-
2 3
mmol/L
c
Glu 0.5 N/A
mmol/L
c
Lac 0.2 N/A
mmol/L
c
Crea 15 15
mol/L
Continued on next page
Analysis Setup
defaults
(continued)
Calibration
Setup defaults
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...