3. Installation and setup
ABL800 FLEX Operator's Manual
Replacement setup,
Continued
Solution
warning
This program allows you to select a solution level value in % below which a
warning will be given.
To select the solution level value for the warning to be activated (default value is
25 %), do the following:
Step Action
1.
Highlight the desired solution bottle: touch the bottle on the screen or
use
Next Bottle
or
Prev. Bottle
.
2.
Select the solution warning level (in %), using
Increase
or
Decrease
.
3.
Select warning values for solution levels in other bottles in the same
manner.
Press
Close
to confirm the settings.
4.
Reagents –
Expiration
Warning
This program allows you to set the time for the analyzer to give a warning before
the calibration and cleaning solutions with additives will expire.
To set the time for the warning, do the following:
Step Action
1.
Select time, using the up/down arrow buttons in the "Give warning
before reagents expire" box.
The actual bottles of the analyzer depend on the analyzer type.
Press
Close
to exit the program and confirm the settings.
2.
Continued on next page
3-52
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...