9. Data management
ABL800 FLEX Operator's Manual
Quality Control Log,
Continued
Viewing a result
To view a result, do the following:
Step Action
1.
Highlight the desired quality control result using scroll facilities.
2.
Press the
Result
button. For details – see
Quality Control Result
in
chapter 5
.
Filter/trend
function
This function allows you to find a quality control result or to select a number of
results in order to make a trend.
To set the filter criteria, do the following:
Step Action
1.
Press the
Filter
button on the
Quality Control Log
screen.
2.
Set the start date by using the 1-day, 7-day, 14-day or 30-day icons in
the “Criteria” box.
Or type in the date using the keypad and confirming the entry with
Enter
.
Change if necessary the end date by highlighting it and typing a date
other than the current date. Confirm with
Enter
.
3.
Use the up and down arrow buttons to highlight the desired search
criteria.
4.
Select the desired choice and confirm each entry with
Enter
.
Or use the keyboard to enter the desired data and confirm each entry
with
Enter
.
Continued on next page
9-12
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...