ABL800 FLEX Operator's Manual
8. Disk functions
Creating a WDC report,
Continued
Storing a WDC
report
(continued)
Step Action
3.
In the “File name box (the
WDC Report
screen), press the keyboard
icon to type the file name. You can change the four characters
“WDC_”.
Confirm the entry on the keyboard and return to the
WDC Report
screen.
4.
Send the file to the selected destination by pressing
Export data
.
Or insert a disk and press
Export data
in the "Export data" box.
Wait until the
WDC Report
screen appears and remove the disk with
the WDC report.
Error messages
The following error messages may appear in the analyzer status:
"Could not create output file" if the wrong disk type (e.g. filled or read-only)
is inserted.
"No statistical data found.
WDC data not generated"
if no data is available for the selected month.