3. Installation and setup
ABL800 FLEX Operator's Manual
Quality control setup,
Continued
QC schedule
(continued)
To remove a quality control measurement from the schedule do the following:
Step Action
1.
Highlight the desired quality control measurement and press
Remove
.
2.
Press the desired button to remove the quality control measurement
from the schedule:
For this day
,
For all days
or
All entries for QC
slot
(number).
Press
Back
to return to the
QC Schedule
screen.
3.
QC Statistics
This program allows you to select the following:
•
the statistics factor
•
the initiation of warning messages at the end of the statistical month.
To select the settings, do the following:
Step Action
1.
Statistics factor expands the control range to the statistics range (it is a
range within which quality control results must fall in order to be
included in the quality control statistics).
Key in the desired statistics factor (from 1.0 to 9.9) on the keypad and
confirm with
Enter
. The default value is 1.5.
Continued on next page
3-42
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...