ABL800 FLEX Operator's Manual
13. Specifications
13-13
Analyzer requirements
Power
Rated voltage: 100-240 V; 50/60 Hz; 270 VA.
Maximum voltage fluctuations:
10 %.
Leakage current measured at 230 V and 50 Hz:
Normal condition (ground connected):
< 0.005 mA
Single fault condition (ground
disconnected):
< 0.5 mA
Fuses
Printer unit has 1 protective fuse: 5 × 20 mm, 4 Amp,
Slow blow (T4AL).
Main fuse includes 2 protective fuses: 5 × 20 mm, 4 Amp,
High break (T4AH). Type Shurter No. 0001.2510.
WARNING – Risk of fire
Fire hazard. Replace fuse as marked.
Start up
Cold start: typically 25 minutes
Warm start: 5 minutes
Gas pressure
Operating pressure: 1.0-38 bar
Maximum pressure: 38 bar
Location Indoor
use
only
Relative humidity
20 to 80 %
Operating
temperature
15
C to 32
C
Storage temperature –20
C to 60
C
Altitude correction
Up to 3000 meters above sea level at standard barometric
pressure of 760 mmHg.
Baro 450-800
mmHg
60.0-106.7 kPa
0.600-1.067 bar
450-800 Torr
Installation category II.
Pollution degree 2.
Analyzer
requirements
Environmental
requirements
Environmental
ratings
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...