ABL800 FLEX Operator's Manual
5. Quality control
AutoCheck measurement
Scheduled
AutoCheck
measurement
A scheduled AutoCheck measurement will start on time provided that no
calibrations are pending. A pending calibration will be performed before the
scheduled AutoCheck measurement.
On the screen that appears when an AutoCheck measurement is about to run
(scheduled either from QC Schedule or from the RADIANCE system), select one
of the three possibilities:
•
No action taken: the AutoCheck measurement starts after 20 seconds (see
counter on the screen)
•
Proceed
pressed: the AutoCheck measurement starts immediately.
•
Postpone
pressed: the AutoCheck measurement will be delayed by five
minutes.
Unscheduled
AutoCheck
measurement
To start an unscheduled AutoCheck measurement, do the following:
Step Action
1.
Press
Analyzer Status – Quality Control
.
2.
Highlight the solution to run an unscheduled measurement on.
3.
Press
Run AC Ampoule
to start the measurement.
The Quality Control Result is shown after the measurement is completed.
NOTICE:
It is only possible to choose ampoules that are packed in the carousel. If
an ampoule is not available, the
Run AC Ampoule
button is inactive.
5-7
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...