ABL800 FLEX Operator's Manual
3. Installation and setup
Quality control setup
Programs
The quality control setup programs are described in this section.
Activate a button to enter the program.
QC solutions
This program allows you to assign or change a quality control solution to a specific
slot.
NOTICE:
Changing a control solution assigned to a slot will delete all current
quality control statistics obtained on that slot. If you want a copy of the statistics
for the last QC month, create a WDC Report – see
chapter 8
:
Disk Functions
.
NOTICE
: (for the ABL8x7 FLEX analyzers only) Slot 11 is assigned to the
Cleaning Met II Solution used for high Crea check.
Continued on next page