3. Installation and setup
ABL800 FLEX Operator's Manual
Quality control setup,
Continued
Westgard Rules
(continued)
Step Action
2.
Press
Next Slot
and activate/deactivate the Westgard Rules for the
other slots in the same manner.
3.
To activate the Westgard rules for specific parameters in a slot,
display the desired slots by using
Next Slot
.
4.
Press the
Edit
button. The screen shows all the Westgard Rules and
their current settings for the parameter.
5.
Display the desired parameter using the
Next Param
or
Prev Param
buttons.
Parameter
6.
Activate the desired Westgard Rule(s) by pressing the corresponding
check button.
Once edited, all future quality control measurements on the parameter
for the given slot will be evaluated according to the selected Westgard
Rule(s).
7.
Select Westgard Rules for other parameters or levels in the same
manner.
8.
To select/deselect all six Westgard Rules for every parameter of a
displayed control solution slot, select the desired slot by pressing the
Next Slot
button.
Continued on next page
3-44
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...