ABL800 FLEX Operator's Manual
11. Troubleshooting
Analyzer messages,
Continued
Messages at
manager level
(continued)
No. Message Interpretation
Operator
action
481-482 Conditioning
error
Indicates a fluid
transport problem of
pH/BG module during
Flush.
484
Today is last
day in
current
statistical
month –
remember to
print QC
statistics
After the current day,
quality control
statistics obtained over
the month will be
deleted and new
statistics started.
Print the QC statistics if a
copy is required.
487 A
new
statistical
month has
begun –
remember to
export WDC
data
A new statistical
month has begun.
Make a WDC report disk.
Removal condition: A
WDC report disk has been
made.
493 Warning:
Bilirubin
detected and
compensat-
ed for
Detected bilirubin
concentration is
greater than
50 umol/L. The
corresponding plasma
bilirubin concentration
can be calculated as
follows: ctBil(blood) =
(1-Hct) x
ctBil(plasma).
No action required.
494 Bilirubin
too
high
Detected bilirubin
concentration,
ctBil(blood), is greater
than 2000umol/L. The
corresponding plasma
bilirubin concentration
can be calculated as
follows: ctBil(blood) =
(1-Hct) x
ctBil(plasma).
No action required.
Continued on next page
11-21
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...