ABL800 FLEX Operator's Manual
2. What Is what
Analyzer status,
Continued
Analyzer Status
elements
(continued)
Button
The colors indicate…
Quality Control
G
REEN
–
OK.
Status
Y
ELLOW
– error(s) in the last QC measurement and/or
QC schedule reminders.
RED –
error-prone parameters are repressed.
Reagents
G
REEN
–
no replacements due at the present time.
Status
Y
ELLOW
– a replacement is due, Calibration/Cleaning
solutions expire soon or Calibration/Cleaning
solutions have expired.
Electrode and
G
REEN
–
no replacements due at the present time.
others
Y
ELLOW
– a replacement is due.
System Messages
G
REEN
–
no (critical) messages.
Y
ELLOW
– non-critical
messages.
R
ED
–
critical messages. The analyzer cannot
calibrate or measure.
AutoCheck
G
REEN
–
no replacements due at the present time.
Status
Y
ELLOW
– a replacement is due.
Calibrations
status
The Calibrations status gives an overview of the status of the most recently
performed calibration of each type and relevant messages.
The following information is available on the screen:
Calibration Type
Lists each calibration type along with its status:
OK
Calibration was accepted.
?
Error(s) detected during calibration.
Pending or overdue calibration. The last
calibration was accepted.
Continued on next page
2-29
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...