ABL800 FLEX Operator's Manual
3. Installation and setup
Quality control setup,
Continued
QC ranges
This program allows you to do the following:
•
globally update all control ranges of a slot to a calculated lot-to-date range.
•
individually edit parameter control ranges by entering your own ranges or
updating to a calculated lot-to-date range.
•
define a minimum allowed control range by entering a Fixed SD (standard
deviation).
See the
Quality Control Systems Reference Manual
for detailed information on
statistics and its parameters.
To edit data for the control solution, do the following:
Step Action
1.
Use
Next Slot
to display the desired slot.
Press
Edit
.
2.
Select the parameter to be edited, using
Next Param.
or
Prev Param.
Selected control
solution
Parameter
Press
Update
to change the range to the one shown in the "Lot-to-Date
range (2 SD)" box - if available (i.e. the range calculated over the
course of the lot, represented mathematically by: mean value
±
2 SD.
This is the range within which 95 % of the measurements taken are
found).
3.
Continued on next page
3-37
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...