11. Troubleshooting
ABL800 FLEX Operator's Manual
Analyzer messages,
Continued
Messages at
manager level
(continued)
No. Message Interpretation
Operator
action
695 ABL
discon-
nected from
RADIANCE
The analyzer was
disconnected from
RADIANCE.
No action required.
696 ABL<>
RADIANCE
communi-
cation error
Communication error
between the analyzer
and RADIANCE.
Contact Radiometer service
representative.
697 Requested
AutoCheck
QC ampoule
not present in
carousel
A scheduled QC
ampoule was absent in
the carousel.
Refill AutoCheck carousel
according to the packing
list.
698 Performing
AutoCheck
on last
ampoule in
schedule
instead of
non-present
If the scheduled QC
ampoule was absent in
the carousel, the
analyzer performs QC
measurement on the
ampoule scheduled
previously.
Refill AutoCheck carousel
according to the packing
list.
699 AutoCheck
QC measure-
ment started
due to cali-
bration error
The analyzer was set up
to perform AutoCheck
measurements in case of
1- or 2-point calibration
errors.
Check Calibration Status
and remedy any reported
calibration errors.
700 Scheduled
AutoCheck
not run due
to errors in
last
calibration
Last calibration
contained an error, and
the analyzer was set up
to suspend AutoCheck
measurements in case
of calibration errors.
Check Calibration Status
and remedy calibration
errors.
703
QC expired
QC measurement is
overdue (corrective
action "Lock analyzer"
has been selected in
Setup program
Corrective Actions).
Perform a quality control
measurement.
Removal condition: No QC
measurements are pending.
704 AutoCheck
QC measure-
ment is
repeated
The scheduled QC
measurement was not
accepted; the
measurement was
repeated as requested in
Setup program
Corrective Actions.
No action required.
Continued on next page
11-38
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...