ABL800 FLEX Operator's Manual
2. What Is what
2-9
FLEXQ module,
Continued
The components are as follows:
Part Function
FLEXQ sampler tray
FLEXQ sampler tray moves to the position for
measurement by means of a step motor (not
shown).
The samplers used in the sampler tray should be
safe
PICO sampler with the
safe
TIPCAP.
Indicator diodes
Change the color from green to yellow when the
optical switch detects the sampler placed in the
slot.
Barcode reader
Laser barcode reader reads the sampler barcode. A
short beep indicates that the barcode has been
read. You can scan the Sampler ID on the
analyzer’s scanner and then place it in the sample
tray or you can place the sampler in the sampler
tray and let the FLEXQ scanner read the sampler
barcode.
Sampler tray cover
The sampler tray cover has three slots for the
samplers; the slots have optical switches for
detecting the sampler. The cover can be removed
(as shown) for cleaning.
Mixer tray
The mixer rotates and moves the ball in the
sampler in order to mix the blood sample. The
sample is adequately mixed after 7 seconds.
The mixer tray moves into position under the
sampler immediately before the measurement.
Parts and
functions
(continued)
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...