3. Installation and setup
ABL800 FLEX Operator's Manual
Analysis setup,
Continued
Reportable
ranges
(continued)
Step Action
5.
To change all parameters to the default values, press the
1-n Set All
Default
button.
Press
Continue
to change all parameters’ reportable ranges to the
default ones.
Press
Cancel
to keep the user-defined reportable ranges and return to
the previous screen.
Press
Close
to exit the program and confirm the selected settings.
6.
See also
Calibration verification
in
chapter 15
in this manual.
Sample Pre-
registration
setup
This program allows you to select interpretation of the barcode and the patient data
that can be confirmed before a sample measurement.
To select the settings, do the following:
Step Action
1.
Use the up/down arrow buttons to select the interpretation of the
barcode setting in the "Interpret barcode input as" box.
Choose one of the following:
•
Patient ID
•
Accession Number
•
Sampler ID.
Continued on next page
3-22
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...