ABL800 FLEX Operator's Manual
3. Installation and setup
Patient reports,
Continued
Editing patient
result layout
(continued)
Step Action
6.
To exclude an item from the selected parameter list, highlight it and
press “
←
”
7.
If you wish to display parameters with their reference ranges and have
them checked against reference ranges/critical limits, do the
following:
•
Highlight a parameter.
•
Press
Show Ranges
to indicate it by “
[
xxx-xxx
]
”.
•
Repeat for other parameters in the same manner.
If the “
[
xxx-xxx
]
” is not included for a parameter, checking against
reference ranges/critical limits is disabled for this parameter.
However, checking against the reportable ranges is always performed.
NOTICE:
Refer to the Reference Manual,
chapter 6: Parameters
, for information
on parameters and their groups.