7. Replacements
ABL800 FLEX Operator's Manual
General information,
Continued
Recording
replacements in
Hold mode
(continued)
Step Action
7.
(conti-
nued)
(In case only the Crea electrodes were remembraned or replaced, the
Confirm Membrane Replacement
screen will be used.)
Press
Accept
to start the restarting sequence or press
Back
to return to
the previous screen.
After the required action is completed, press
Restart
and then
Accept
.
Entering Fluid
Replacements
mode
The Fluid Replacements mode ensures that the proper restart sequence will follow
the exit from this mode.
To enter the Fluid Replacements mode, do one of the following:
•
Press
Analyzer Status – Reagents – Replace
.
•
Remove a solution container
Recording
replacements in
the Fluid
Replacements
mode
To record replacements in the Fluid Replacements mode, do the following:
Step Action
1.
Press
Analyzer Status – Reagents – Replace.
2.
Remove the component to be replaced according to recommended
procedure.
Continued on next page
7-4
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...