ABL800 FLEX Operator's Manual
13. Specifications
13-5
Input parameters
All the input parameters available on the ABL800 FLEX analyzers, independent of
configuration, are listed below.
Symbol Parameter
Unit
Input
range
Defaults
T
Patient
temperature
o
C 15.0-45.0
37
o
F
59.0-113.0
98.6
QUA/Au
o
C 18.0-32.0
25
ampoule temperature
o
F
64.4-89.6
77
F
O
2
(I) Fraction of oxygen in
%
0.0-100.0
21.0
dry inspired air
fraction 0.000-1.000 0.210
c
tHb
Total hemoglobin concentration
g/dL
0.0-33.0
-
(if not measured)
g/L
0.0-330
-
mmol/L
0.0-20.5
-
RQ
Respiratory quotient – ratio of
carbon dioxide production and
oxygen consumption
- 0.00-2.00
0.86
p
O
2
(v
–
) Oxygen tension in mixed
mmHg
0.0-xxx.x*
-
venous blood
kPa
0.00-xx.xx*
-
s
O
2
(v
–
) Oxygen saturation
%
0.0-100.0
-
in mixed venous blood
fraction 0.000-1.000 -
Q
·
t
Cardiac output
L/min
0.0-xxx.x*
-
V
·
O
2
Total oxygen consumption
mL/min
0-xxxx*
-
mmol/min 0.0-xxx.x*
-
V
CO
Volume of carbon monoxide
mL
0.0-xxx.x*
-
p
50(st) Half saturation tension of oxygen
mmHg
0.01-100.00
26.84
in blood at standard conditions
kPa
0.001-13.332
3.578
F
COHb
(1)
The fraction of COHb measured
before the CO-injection
%
fraction
0.0 -100.0
0.000-1.000
-
-
F
COHb
(2)
The fraction of COHb measured
after the CO-injection
%
fraction
0.0 -100.0
0.000-1.000
-
-
* numerical format
List of
parameters
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...