11. Troubleshooting
ABL800 FLEX Operator's Manual
Analyzer messages,
Continued
Messages at
manager level
(continued)
No. Message Interpretation
Operator
action
966 Cleaning
–
Additive
will expire
soon
The Cleaning additive
will expire soon.
Replace the solution before
the lifetime is actually
exceeded.
967 Creatine
high.
Unable to
perform
Crea
measure-
ment
The Creatine
concentration is too
high compared to the
Crea concentration and
the analyzer is not able
to measure Crea within
the specifications.
See chapter 14:
Specifications in the
Operator's Manual
- Check for and remedy
other errors related to the
result, system messages or
calibration status.
- Perform a High Crea
Check. If the High Crea
Check result is accepted,
the blood sample may be
suspected.
- Perform measurement on
a new blood sample.
- Perform the Electrode
Troubleshooting procedure
969 FLEXQ
module not
expected
Software configuration
does not match
analyzer type.
Contact Radiometer
Service Representative.
Removal condition:
Matching software and
hardware configuration.
970 Rilibäk
violation:
Value above
upper limit
971 Rilibäk
violation:
Value belov
lower limit
Continued on next page
11-70
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...