ABL800 FLEX Operator's Manual
3. Installation and setup
Replacement setup,
Continued
Replacement
schedule
(continued)
Step Action
2.
Change the interval for the selected replacement action (displayed in
the "Action" box) with the up/down arrow buttons (see recommended
replacement intervals below).
NOTICE:
The replacement schedule reminder "Lock analyzer when
10 % overdue" (selected in Corrective Actions – see further in this
chapter) will work on the basis of the setting selected in the "Action"
box.
3.
Touch to highlight the "Next date" box and change the date using
screen keypad. Confirm with
Enter
.
4.
Press
Back
to return to the
Replacement Setup
screen.
Repeat steps
1-4
for each replacement action to be scheduled.
5.
The following replacement intervals are recommended by Radiometer.
NOTICE:
The replacement intervals below are guidelines only. Under no
circumstances do they guarantee the lifetime of the replacement items. For
analyzers with high throughput the replacement intervals should be adjusted
accordingly in the Replacement schedule.
Recommended
replacement
intervals
For an average of 40 samples per day
:
Action Interval
Replacing Crea A and Crea B electrode membranes
14 days
Replacing metabolite membranes
1 month
Replacing blood gas and electrolyte electrode membranes
3 months
Replacing pump tubing
6 months
Replacing inlet gasket
6 months
Replacing fan filter
as required
Replacing electrodes
as required
Replacing reference electrode membranes:
40 samples per day:
All analyzers, except ABL8x7 FLEX analyzers
1 month
ABL8x7 FLEX analyzers only
14 days
≤
40 samples per day:
>40 samples per day:
ABL8x7 FLEX analyzers only
1 week
The status of solutions, waste container and gas cylinders are not scheduled and
can be viewed in the Replacements on the
Analyzer Status
screen.
Continued on next page
3-49
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...