7. Replacements
ABL800 FLEX Operator's Manual
Replacing membranes or electrodes,
Continued
Replacing an
electrode
(continued)
Step Action
5.
All electrodes:
Dry the electrode contact and place the new electrode in the
measuring chamber.
6.
If no more replacement actions are to be made, mount all covers and
parts, and press
Restart
.