ABL800 FLEX Operator's Manual
3. Installation and setup
Quality control setup,
Continued
QC Statistics
(continued)
Step Action
2.
To set up the analyzer to send reminders to print or export the monthly
statistics, activate the appropriate check button (or both) for the
message(s) to appear in System Messages upon occurrence.
Press
Close
to exit the program.
3.
Westgard Rules
The Westgard Rules are a set of statistical rules that, when applied to the quality
control results, can increase the probability of detecting an error in the sampling
procedure or in the analyzer itself, or help detecting a shift or trend in your quality
control results by comparing current measurement values of a control solution with
previous values.
This program allows you to select Westgard Rules for all slots or for specific
parameters.
To activate or deactivate the assigned Westgard Rules, do the following.
Step Action
Press
On/Off
button to activate the assigned Westgard Rules for the
slot.
1.
Slot, solution,
type, lot
Press
On/Off
again to deactivate the assigned Westgard Rules.
Buttons are
grayed out
Continued on next page
3-43
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...