6. Calibration
ABL800 FLEX Operator's Manual
Unscheduled calibrations
Calibrations called by the operator at any time are called
unscheduled
calibrations.
Calling an
unscheduled
calibration
To call an unscheduled calibration, do the following:
Step Action
1.
Check that the analyzer is in the Ready mode.
2.
Press
Menu
–
Start Programs
–
Calibration Programs
.
Call a calibration.
If a button is grayed out, a calibration cannot be called (as 1-point
calibration on the screen).
3.
To see the electrode updatings during a calibration, go to the
Analyzer Status
screen and press the
Electrode Upd
. button.
Calibrations
during 24 hours
after restart
The 24-hour calibration period is only necessary after the replacement of
electrodes and/or electrode membranes.
The calibration intervals for the first 24 hours after restart
are as follows:
Time after restart
Calibration intervals
First 4 hours
1-point calibration
Every 30 minutes
2-point
calibration
Every
hour
Remaining 20 hours
1-point calibration
Every hour
2-point calibration
Every 4 hours
NOTICE:
If the restart is caused by
c
Glu/
c
Lac electrode/membrane maintenance
1-point calibrations will be performed every 30 minutes for the first 4 hours and
every hour for the following 20 hours. 2-point calibrations will follow the normal
schedule.
NOTICE:
If the calibration schedule is set up to perform calibrations at intervals
less frequent than those programmed for the 24-hour calibration period, the
calibration schedule is overridden.
6-4
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...