3. Installation and setup
ABL800 FLEX Operator's Manual
Analyzer security,
Continued
Operators and
Passwords
This program allows you to add, edit or remove operators and to assign an access
profile to each operator.
NOTICE:
If the Centralized User Management option is enabled in the
General
Security
screen, you cannot add, remove or edit the operators, but only view the
access profiles of the individual operators.
To enter the Operators and Passwords program, press
Menu – Utilities – Setup –
Analyzer Security – Operators and Passwords
.
Shows the list of
actual operators
Shows the list of
available operator
profiles
When the analyzer is taken into use, the following default operators are available:
Operator
Has access to…
Manager
All menu items and programs (not service programs). It is
recommended to remove this operator with the standard
password: 123456, and enter the actual users with their
profiles and passwords.
Radiometer
All menu items and programs (user and service) on the
analyzer. Note that "Radiometer" cannot be removed from
the operator list.
Remote operator
Only if the Remote Support option key is installed.
Continued on next page
3-8
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...