ABL800 FLEX Operator's Manual
11. Troubleshooting
Electrode troubleshooting procedures,
Continued
To troubleshoot the
c
Glu and
c
Lac electrodes, do the following:
cGlu and cLac
troubleshooting
procedure
Step Action
1.
Remove the electrode and check the measuring chamber for clots;
clean the measuring chamber with a cotton stick moistened with
distilled water. Make sure that no cotton fibers are left in the
measuring chamber.
2.
Replace the electrode membrane.
3.
Call a 1-point calibration. If the error persists, replace the electrode.
To troubleshoot the
c
Crea-A and
c
Crea-B electrodes, do the following:
cCrea electrodes
troubleshooting
procedure
Step Action
1.
Remove the electrodes and check the measuring chamber for clots;
clean the measuring chamber with a cotton stick moistened with
distilled water. Make sure that no cotton fibers are left in the
measuring chamber.
2.
Replace the membranes of both electrodes.
3.
Call a 2-point calibration. If the error persists, replace the electrodes.
Refer to
chapter 7
for specific instructions on replacing electrodes and membranes.
Remember to record all replacements.
NOTICE:
11-79
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...