3. Installation and setup
ABL800 FLEX Operator's Manual
Analysis setup,
Continued
Selecting HbF
correction
To give the user the possibility of selecting or deselecting HbF correction during a
measurement, do the following:
Step Action
1.
Select HbF correction for all levels or for HbF levels higher than 20 %
in Miscellaneous Setup (
Setup
−
General Setup
−
Miscellaneous
Setup
) – see the description further in this chapter.
2.
Return to the Analysis Setup, choose the desired syringe or capillary
program and press the
Parameters
button to enter the
Parameter
Profile
screen.
3.
Activate the check button in the "Use HbF correction" box (if
"Disabled for all levels" has been selected in Miscellaneous Setup, the
text box and the check button in it will not be visible on this screen).
4.
Press the dynamic parameters check button in the “Use dynamic
parameters” box in order to be able to exclude the HbF correction and
other parameters during a sample measurement.
HbF correction
was disabled in
Miscellaneous
Setup
Continued on next page
3-16
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...