11. Troubleshooting
ABL800 FLEX Operator's Manual
Pump troubleshooting procedure
Procedure
To troubleshoot pumps, do the following:
Step Action
1.
Check the Rinse Solution container for sufficient volume and proper
mounting. Replace as required.
2.
Inspect the pump tube(s) for leaks, wear and proper seating.
Replace as required.
3.
Call the Liquid Sensor Adjustment program.
4.
Call the Pump Calibration program.
This program is performed automatically after any replacements
made in step
2
above.
5.
If the error persists, contact a service representative.
11-80