1. Introduction
ABL800 FLEX Operator's Manual
Limitations of use and known interfering substances,
Continued
Substance Interference
Carboxymethyl cellulose (CMC)
Some auto-venting arterial blood samplers
contain carboxymethyl cellulose (CMC) in
the porous vent. CMC can dissolve into the
sample and give erroneously low
c
Ca
2+
results. Therefore we recommend
Radiometer accessories together with our
analyzers, e.g. the
safe
PICO arterial blood
sampler which is specifically designed to
minimize sample contamination with CMC.
Galactose, glucosamine, maltose,
mannose, xylose
Erroneously high
c
Glu measurements.
CAUTION -
Known
interfering
substances
(continued)
For detailed information – see
Interference Tests
in
chapter 5
of the
ABL800 FLEX
Reference Manual
.
1-6
Содержание ABL800 FLEX