ABL800 FLEX Operator's Manual
6. Calibration
tHb calibration,
Continued
tHb calibration
check
To verify the validity of the tHb calibration, do the following.
Step Action
1.
Open the syringe inlet flap and place the adapter with the S7770
solution in the inlet.
2.
Select the mode:
Syringe – S195
μ
L
or
Syringe – S85
μ
L
or
Syringe – S250 µL
depending on the ABL800 FLEX version.
3.
Press the
Start
button. Remove the ampoule when prompted and close
the inlet.
4a.
If the calibration check value* is within the insert limits, the tHb
calibration is accepted.
4b.
If the calibration check value* is outside the insert limits, repeat the
tHb calibration and then the calibration check immediately after it.
If the second tHb calibration and the calibration check fail again,
contact a service technician.
* Can be
c
tHb, or
c
tHb and
c
tBil
,
or ctBil.
NOTICE:
If the tHb calibration check has not been performed immediately after the tHb
calibration, repeat the calibration and the calibration check immediately after it.
6-7
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...