3. Installation and setup
Quality control setup,
Continued
QC schedule
(continued)
Step Action
3.
Select the slot/quality control solution (entered in the Quality Control
Solutions program), using the up/down arrow buttons in the box.
Confirm with
Select
.
4.
Highlight the "Weekdays" box using the
Field down
button.
Activate the relevant check buttons to select the days of the week this
measurement should be performed.
5.
Highlight the "Start time" box.
Key in the time to perform a measurement and confirm with
Enter
on
the keypad.
NOTICE:
The symbols for the automatic or manual quality control
(given on the previous page) will automatically appear in the
schedule.
6.
Highlight the "Repeat" box. Select the interval the measurement
should be repeated, using the up/down arrow buttons in the box.
7.
Press
Back
to return to the
Quality Control Schedule Setup
screen.
If more than one quality control solution was scheduled, the schedule
screen will look like this:
To edit the schedule, press
Edit
and follow the procedure above.
Continued on next page
3-41
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...