ABL800 FLEX Operator's Manual
11. Troubleshooting
Analyzer messages,
Continued
Messages at
manager level
(continued)
No. Message Interpretation
Operator
action
955 Measured
QC value
lower than
statistical
range
(Crea A)
The parameter value is
below the lower limit
of the user-defined
statistical range.
Measurement is not
included in the
statistics.
- Verify procedure and
repeat the measurement.
- Refer to the Quality
Control Systems Reference
Manual for details on the
evaluation of the results.
956 Measured
QC value
higher than
statistical
range
(Crea A)
The parameter value is
above the upper limit
of the user-defined
statistical range.
Measurement is not
included in the
statistics.
- Verify procedure and
repeat the measurement.
- Refer to the Quality
Control Systems Reference
Manual for details on the
evaluation of the results.
958 Solution
lifetime
exceeded
A solution is more than
14 days old and cannot
be used for Crea
calibration or High
Crea Check.
Calibration Solutions:
- Crea is no longer
reported during
measurements and is
marked red on the
parameter bar.
Cleaning Solution:
- It is no longer
possible to perform a
High Crea Check.
Replace required solution
bottle - remember to add
the additive.
Removal condition:
Fresh bottle installed.
Continued on next page
11-67
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...