ABL800 FLEX Operator's Manual
11. Troubleshooting
Analyzer messages,
Continued
Messages at
manager level
(continued)
No. Message Interpretation
Operator
action
671 Analyzing
Unit service
setup non-
default
An option in the
service setup was
changed from default
setting.
Contact Radiometer service
representative.
Removal condition:
Options restored to default
values.
677 Unreliable
A
measurement
or
calibration was
attempted during
"Conditioning error in
El/Met electrode
module".
- Check whether
Conditioning error (525)
was reported during the last
analyzer activity. If not,
repeat measurement/
calibration.
- Check Rinse Solution
level in the container.
- Perform Rinse.
Removal condition: Rinse
without Conditioning error.
678
Heater error One or more measuring
modules are over-
heated.
- Ensure the ambient
temperature is between 15
and 32 degrees Celsius.
- Shut down the analyzer
using the Temporary
Shutdown function, then
restart the analyzer.
- Replace fan filter if dirty.
- Shield analyzer from
direct sunlight and other
heat sources.
- Contact Radiometer
Service Representative.
Removal condition:
Analyzer temperature
within required range.
679 Barometer
error
Measured parameter
may be unreliable due
to barometer error.
Contact Radiometer service
representative.
Continued on next page
11-35
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...