ABL800 FLEX Operator's Manual
4. Sample measurements
General information,
Continued
5.
CAUTION - Risk of incorrect result
When measurements are flagged always conduct the operator
actions required by the troubleshooting program to prevent possible
clots from remaining in the sample path. Fluid path clots may cause
erroneous measurement results.
6.
CAUTION - Risk of erroneous results
Always meticulously follow the sampling procedures described in
chapter 12: Sampling
. Failure to follow these procedures may
introduce clots or air bubbles in the sample and yield erroneous
results.
CAUTION
notices
(continued)
When the FLEXQ is in use, the
CAUTION
notices 2 and 4 above do not apply.
Interrupting a
measurement
To interrupt a measurement, press
(Stop) button.
Before using the
FLEXQ
To use the FLEXQ efficiently, the setup should be made thoroughly.
Program Recommendations
Analysis Setup –
Syringe Modes
•
Make the report layout for each measuring program
you are going to use, including the Patient ID items
and parameter profile
•
Select the default layout for each selected measuring
program.
This will define the number of measuring programs to
select from before each measurement.
Analysis Setup –
Sample Pre-
registration
Select interpretation of the barcode entry.
Analysis Setup –
Sample Logistics
•
Select batch mode for automatic measuring process or
deselect it if single measurements are to be performed
•
Select sample age for each measured parameter
Continued on next page
4-7
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...