ABL800 FLEX Operator's Manual
11. Troubleshooting
Analyzer messages,
Continued
Messages at
manager level
(continued)
No. Message Interpretation
Operator
action
876 FLEXQ
positioning
warning
A problem was
detected during
positioning of the
sampler tray
The message is informative
and intended for service
use only
Removal condition:
No FLEXQ positioning
problems observed for 24
hours.
877 FLEXQ
positioning
error
A sampler in the
sampler tray could not
be aligned with the
syringe inlet; aspiration
of the sample was not
possible.
- Remove all the samplers
from the sampler tray (the
sampler registration is
lost).
- If the sampler tray is
blocking the syringe inlet
or AutoCheck module, try
to push it gently to its
home position.
- If the syringe inlet is not
blocked, scan the sampler
barcode, using the
analyzer’s barcode reader
and perform a manual
measurement.
- To get the FLEXQ
operational again, restart
the analyzer, using
Temporary Shutdown and
power off/on.
- If the error persists,
contact Radiometer service
representative.
- Removal condition:
Successful FLEXQ
sampler tray calibration.
Continued on next page
11-55
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...