ABL800 FLEX Operator's Manual
5. Quality control
Quality control result,
Continued
High Crea
Check
(for the ABL8x7
FLEX analyzers
only) (continued)
Crea
diff.
is the difference between the expected creatinine value in the Cleaning Met
II Solution and the measured value. It is shown on the plot – see page 5-14.
To evaluate the Crea
diff.
accurately, it is required to enter the expected ambient
temperature for the 14-day period in the Environment Setup – see
Analyzer settings
in
chapter 3.
The ambient temperature on the
High Crea Check ID
screen is grayed out, i.e. it
can only be viewed. The ambient temperature setting can be changed in the
Environment Setup.
The
Quality Control Plot
screen appears once the
Plot
button is pressed.
Quality control
plot
The following diagram shows the details of a typical plot.
Time
Measurement
Slot number
Solution type
Out-of-range symbol
Control range upper limit
Control range lower limit
Measured value
Date
Lot change
Parameter
Continued on next page
5-13
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...