ABL800 FLEX Operator's Manual
7. Replacements
Replacing membranes or electrodes,
Continued
Remembraning
an electrode
(continued)
Step Action
4.
All electrodes:
Press the electrode firmly into the jacket until a click is heard.
Further for the reference electrode:
Check that the old O-ring has been removed and press the electrode
through the protective film covering the electrode jacket.
CAUTION – Safety precautions
Solution is irritating to eyes, respiratory system and skin. In
case of contact with eyes, rinse immediately with plenty of
water and seek medical advice.
5.
All electrodes:
Remove the membraned electrode from the membrane box.
Further for the reference electrode:
Rinse the electrode with tap water and dry with a lint-free tissue.
Further for
p
CO
2
and
p
O
2
electrodes:
Remove any air bubbles trapped between the electrode tip and the
membrane by pressing the tabs on the electrode jacket and moving the
electrode up and down slightly in the electrolyte until the bubble
disappears.
6.
All electrodes:
Dry the electrode contact.
7.
All electrodes:
Place the electrode into the measuring chamber and close the
electrode connector, pushing on its top until it clicks shut.
8.
If no more replacement actions are to be made, mount all covers and
parts, and press
Restart
.
Replacing an
electrode
To replace an electrode, do the following:
Step Action
1.
All electrodes:
Remove the old electrode from the measuring chamber.
2.
For the new pH,
p
CO
2
and
p
O
2
electrodes only:
2a.
Rinse the electrode under tap water.
2b.
Shake it to remove excess drops of water.
Do not dry the
electrode
.
4.
All electrodes, except the pH electrode:
Membrane the electrode at once.
Continued on next page
7-7
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...