ABL800 FLEX Operator's Manual
11. Troubleshooting
Analyzer messages,
Continued
Messages at
manager level
(continued)
No. Message Interpretation
Operator
action
605 Cal
expired
606 Cal
expired
(pH)
608 Cal
expired
(
p
CO
2
)
609 Cal
expired
(
p
O
2
)
610 Cal
expired
(K)
611 Cal
expired
(Na)
612 Cal
expired
(Ca)
613 Cal
expired
(Cl)
614 Cal
expired
(Glu)
615
Cal expired
(Lac)
616 Cal
expired
(OXI)
Too long time passed
since the last successful
calibration of the
parameter. Calibrations
of the parameter are
considered invalid and
values reported as
".....".
Perform 1- or 2-point
calibration.
Removal condition:
Successful 1- or 2-point
calibration.
618 Waste
detector
error
The waste detector may
be faulty or damaged.
Contact Radiometer service
representative.
Removal condition:
Functioning waste detector
installed.
619 Waste
container
missing
The waste container is
missing or improperly
mounted.
Mount the waste container
properly.
Removal condition: Waste
container properly
mounted.
Continued on next page
11-29
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...