ABL800 FLEX Operator's Manual
3. Installation and setup
Miscellaneous setup,
Continued
List of options
(continued)
Option Function
Log all
measurement
activities
If enabled, "Ready", "Rinse", "Aspirating" and
"Measurement" will be registered in the Activity log.
Otherwise these activities will not be registered in the
Activity log. This option aims to avoid too many entries
in the Activity log.
Auto temp. unit
conversion
°
C will be automatically changed to
°
F if the entered
temperature is over the value of 45.
Show screen saver
The screen saver will show up if the analyzer was idle for
10 minutes.
Show parameter bar If disabled, the parameter bar will not be shown on the
Ready
screen.
To activate/deactivate an option, do the following:
Step Action
1.
Scroll the list of options with the up/down arrow keys.
2.
Highlight the option and press the check button (
3
) beside it. To
deactivate the option, press the check button again.
Press
Close
to confirm the settings and return to the
Ready
screen.
3.
Selecting HbF
correction
option
This option disables HbF correction for all levels, or enables it for all levels or for
HbF levels higher than 20 % (HbF correction will be automatically disabled if
F
COHb <
−
0.5 % in order to decrease a false HbF detection on adult blood
samples and to ensure non-biased oximetry results).
The guidelines for selecting/deselecting HbF correction are as follows:
For neonatal samples: Use "Enabled for all levels".
It is important to enable HbF correction in order to
obtain correct results for
c
tBil,
s
O
2
,
F
O
2
Hb,
F
MetHb,
F
COHb and
F
HHb.
For adult samples:
Use "Disabled" or "Enabled for levels > 20 %".
NOTICE:
When an adult sample is measured with HbF correction "Enabled for all
levels" or "Enabled for levels > 20 %", it will slightly affect measurement of
s
O
2
,
F
O
2
Hb,
F
MetHb,
F
COHb and
F
HHb, and will cause a marginal number of adult
samples reported with HbF present.
If this is not acceptable, the HbF correction for adult samples can be turned off
before a measurement – see
Selecting HbF Correction
in the Analysis Setup in this
chapter.
To select the desired option, use the arrow buttons in the box.
Continued on next page
3-87
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...