3. Installation and setup
Quality control setup,
Continued
Rilibäk Ranges
(continued)
Step Action
7.
To select the d/- range press the desired radio button.
8.
Enter the d/- range in the "Ranges" box.
Press
Back
to return to the
Rilibäk Ranges
screen. The added Rilibäk
rule is now shown in the screen.
9.
To edit a Rilibäk rule, do the following:
Step Action
1.
Select the desired rule in the
Rilibäk Ranges
screen and press
Edit
to
display the screen below:
2.
Use the
Right
or
Left
arrows to jump between the input boxes and edit
the desired values.
Press
Back
to return to the
Rilibäk Ranges
screen.
3.
To remove a Rilibäk rule, do the following:
Step Action
Highlight the desired rule in the
Rilibäk Ranges
screen and press
Delete
.
1.
NOTICE:
When a quality control measurement violates an applied Rilibäk Rule, a
red R is shown in front of the parameter name in the result.
3-47
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...