3. Installation and setup
ABL800 FLEX Operator's Manual
Communications,
Continued
LIS/HIS
Connection
Setup
(continued)
Step Action
2.
Press
Keyboard,
type the name of the connection instead of the default
one, and press
Enter
.
Press
Back
to return to
LIS/HIS Connection Setup
screen.
3.
Select the high-level protocol according to the requirements of the
connected device.
Select between: ASTM, ASTM6xx, HL7 version 2.2, HL7 version 2.5
or POCT1-A DML, using the up/down arrows in the box.
4.
Select the low-level protocol as follows:
−
Use “Serial” or “Serial (RAW)” for the serial connection.
−
Use “Network (TCP/IP)” for the network connection.
−
Use “Network (TCP/IP)ASTM" for additional serial connection (not
all combinations of high-level protocols and low-level protocols are
possible).
Continued on next page
3-70