ABL800 FLEX Operator's Manual
6. Calibration
General information,
Continued
NOTICE:
Additional 1-point calibrations are performed on the analyzers with the metabolite
electrodes:
•
after 11 measurements made one after another, no more than 10 minutes apart
and
•
30 minutes after the last measurement was made.
The additional 1-point calibration does not change the intervals selected for the
scheduled calibrations.
In cases when a scheduled 2-point calibration is imminent, the additional 1-point
calibration will be substituted with the 2-point calibration.
NOTICE:
(for the USA only) If an unacceptable pH drift is detected during a 1 point pH/BG
calibration, the ongoing calibration is aborted and a normal 1-point calibration is
performed.