3. Installation and setup
Replacement setup,
Continued
Solution Level
Forecast
This program allows you to change the solution level forecast so that the analyzer
will predict the solution levels in the bottles more precisely.
It will improve both the graphical presentation on the
Replacement Status
screen
and the accuracy of the solution low level warning.
To change solution level forecast, do the following:
Step Action
1.
Select the solution bottle, using the
Next bottle
or
Prev. bottle
button
(the Rinse solution has been selected on the screen).
2.
Select the true level (in %) with the up and down arrow buttons on the
screen.
The bottom of each solution bottle has markings for solution levels.
Other Solutions
Rinse Solution
You can also remove the solution bottle to check the level, and the
analyzer will not enter Hold mode. However, the Refill will be
performed as soon as the bottle has been mounted back again.
25 %
min.
25 %
min
.
The selected setting will be displayed on the
Analyzer Status
screen.
The screen below is an example.
Continued on next page
3-53
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...