ABL800 FLEX Operator's Manual
4. Sample measurements
Patient result,
Continued
Step
Patient result
approval
(continued)
Action
Reject
Press
Accept
to reject the result and to send it to the
connected RADIANCE system or LIS/HIS.
Press the keyboard icon to write a note.
Press
Cancel
to return to the previous screen.
Rerun
Press
Accept
to mark that the measurement should be
repeated, and to send it to the connected RADIANCE
system or LIS/HIS.
Press the keyboard icon to write a note.
Press
Cancel
to return to the previous screen.
Note
Press the keyboard icon to type a note prior to the
approval or to edit a highlighted note.
To delete a highlighted note, press the keyboard icon and
delete the note.
Once the result has been approved/rejected/rerun, the ID data can only
be viewed, not changed (grayed-out input boxes). The
Parameters
,
Patient Lookup
and
Request
buttons will not be available either.
In the Patient Report log you can filter patient reports according to the
approval status.
The
Approval
button changes to
Send
(if an external connection is
used) after the approval of the patient result has been made.
If result approval is used, data will not be transmitted to a LIS/HIS
connection before the approval has been made.
Press
Acid-Base Chart
to view the acid-base chart for the selected patient result
(the button will not be available if pH and
p
CO
2
were deselected in the Parameter
setup. The sample type should be either “Arterial” or “Capillary”).
Acid-Base Chart
The chart illustrates the patient result according to the Siggard-Andersen
nomogram. The clinical conditions corresponding to each of the 8 zones listed to
the right of the screen.
Continued on next page
4-23
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...