3. Installation and setup
ABL800 FLEX Operator's Manual
Parameters and input setup,
Continued
Parameters can be
locked/unlocked
from the RADIANCE system.
To lock
or unlock
a parameter from the analyzer, do the following:
Parameters
(continued)
Step Action
1.
Make sure the analyzer is not connected to the RADIANCE system.
Highlight a parameter on the screen, using the scroll facilities.
Press
Lock/Unlock
. (This button is grayed out if the analyzer is
connected to the RADIANCE system.)
The locked parameter will show red on the parameter bar and will
change the overall analyzer status traffic light on the
Analyzer Status
button to red. The parameter value will be absent from the printout.
However, the locked parameter will be calibrated.
2.
To unlock a parameter, highlight it and press
Lock/Unlock
. The traffic
light on the
Analyzer Status
button will change from red to a color
corresponding to the analyzer’s overall status.
3.
To
edit
the parameter setup, do the following.
Step Action
1.
Highlight the desired parameter and press
Edit
.
2.
Activate (or deactivate) the following check buttons to select (or
deselect) the following functions:
Highlighted
parameter
Continued on next page
3-56
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...