11. Troubleshooting
ABL800 FLEX Operator's Manual
Electrode troubleshooting procedures
Using the
procedures
The actions are listed in order of priority. It is recommended to perform each
action and verify whether or not the error has been remedied. If not, proceed to the
next step.
pH, cK
+
, cNa
+
,
cCa
2+
, and cCl
–
troubleshooting
procedure
To troubleshoot pH,
c
K
+
,
c
Na
+
,
c
Ca
2+
, and
c
Cl
–
electrodes, do the following:
Step Action
1.
The pH electrode only:
Clean the electrode, using a cotton swab moistened with Cleaning
Solution.
2.
Check the measuring chamber for clots: clean the measuring chamber
with a cotton stick moistened with distilled water. Make sure that no
cotton fibers are left in the measuring chamber.
3.
Replace the electrode membrane (not the pH electrode).
4.
Call a 2-point calibration. If the error persists, replace the electrode.
If an error has occurred on two or more parameters, troubleshoot the reference
electrode.
Reference
electrode
troubleshooting
procedure
To troubleshoot the reference electrode, do the following:
Step Action
1.
Remove the electrode and check the measuring chamber for clots:
clean the measuring chamber with a cotton stick moistened with
distilled water. Make sure that no cotton fibers are left in the
measuring chamber.
2.
Replace the electrode membrane.
3.
Call a 2-point calibration. If the error persists, replace the electrode.
To troubleshoot the
p
CO
2
and
p
O
2
electrodes, do the following:
pCO
2
and pO
2
troubleshooting
procedure
Step Action
1.
Remove the electrode and check for air bubble(s) trapped between the
electrode tip and the membrane.
If any air bubbles exist, remove them by pressing the tabs on the
electrode jacket and moving the electrode slightly up and down until
they dissipate.
2.
Replace the electrode membrane.
The
p
O
2
electrode: brush the electrode tip with the supplied brush
before remembraning the electrode.
3.
Call a 2-point calibration. If the error persists, replace the electrode.
Continued on next page
11-78
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...