11. Troubleshooting
ABL800 FLEX Operator's Manual
Analyzer messages,
Continued
Messages at
manager level
(continued)
No. Message Interpretation
Operator
action
475 No
sample
in Oxi
liquid
sensor
within time
limit
The sample did not fill
the oximetry module
properly.
- Check the sample for
clots.
- Ensure sample volume
matches the selected
sample mode.
- Perform the Fluid
Transport Troubleshooting
procedure.
476 Measure-
ment
unstable
The electrode response
fault occurred during
measurement.
- Check for and remedy
other errors in System
Message or Calibration
Status.
- Repeat the measurement.
- Perform any pending
replacements, including
electrodes.
- Ensure the electrode is
properly installed.
- Ensure the analyzer is
properly grounded.
- Perform QC. If the QC
result value is within the
control range, the blood
sample may be suspected.
Continue troubleshooting.
- Perform the Electrode
Troubleshooting procedure.
478
p
CO
2
drifting
The
p
CO
2
drift value
exceeds the user-
defined 0.3
kPa during flush.
480
p
O
2
drifting The
p
O
2
drift value
exceeds the user
defined 0.3
kPa during flush.
- Call 1-point calibration.
- Check for and remedy
any specific errors in the
last calibration result, then
call another 1-point
calibration.
- Contact Radiometer
service representative.
Continued on next page
11-20
Содержание ABL800 FLEX
Страница 1: ...ABL800 FLEX operator s manual ...
Страница 2: ......
Страница 3: ...ABL800 FLEX operator s manual ...
Страница 4: ......
Страница 12: ......
Страница 18: ...Contents ABL800 FLEX Operator s Manual ...
Страница 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Страница 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Страница 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Страница 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Страница 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Страница 436: ......
Страница 438: ...Date of issue ABL800 FLEX Operator s manual ...