Setting Injector Temperatures . 71
Oven Temperature Programming Mode
In this mode the POC or PSS injector is configured as Oven. The injector temperature is set to the oven
temperature plus 5
°
C. This is the default mode and the easiest mode to operate in. For example, if the
oven program is as follows:
Parameter
Value
Temp 1
50 °C
Time 1
2 min
Rate 1
5 °C/min
Temp 2
240 °C
Time 2
10 min
The initial injector temperature will be 55
°
C. When the GC run begins, the injector will hold this
temperature for 2 minutes, then it will ramp up to 245
°
C at 5
°
C/min and hold for 10 min.
NOTE: Opening the oven door turns off the oven and the injector heater. Also, if you reset the oven temperature
during the run to the next level, the injector temperature will follow.
Inlet Temperature Programming Mode
The inlet temperature programming mode is used when the POC or PSS injector is configured as Inlet. In
this mode, the injector follows the injector temperature program that you entered in the method.
The initial default values are: Initial Temp 1 is 50
°
C, Injection Time is
999 min, and Inj Rate is 0. With nonprogrammed injectors, an entry of 0
turns the heated zone off. With programmed injectors, a temperature of
0 is possible. Therefore, when
0
is entered for the initial injector
setpoint, the following screen appears:
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...