Connecting the PPC Modules . 231
Connecting the PPC Modules
The Clarus GC PPC modules have been properly installed and configured at the factory for your specific
combination of injectors and detectors. The configuration information is stored in a reserved area of the
battery-backed memory; however, if the memory lost the PPC module connect configuration or if you
installed a new PPC module by purchasing an add-on kit, you must follow the procedure(s) listed in this
section.
NOTE: By connecting a PCC module, you actually set the Clarus GC controlling software to recognize a specific
PPC module that is installed at a specific location. This is necessary so that the software can control the
installed hardware.
The PPC board has 12 sockets to which various injector and detector modules are connected. The socket
locations and descriptions are listed below and shown in the following table:
Injector and Detector Socket Descriptions
Injector 1
1 (J10)
Injector 2
2 (J11)
Injectors (Carriers)
Detector 1
Air or TCD Ref
3 (J12)
Detector 1
HYD or Make
Up
4 (J13)
Detector 1
Detector 2
Air or TCD Ref
5 (J14)
Detector 2
HYD or Make
Up
6 (J15)
Detector 2
7 (J16)
8 (J17)
Aux
9 (J18)
10 (J19)
11 (J20)
12 (J21)
Split Vent for CAP & PSS injectors or
Pressure Readout for PKD & POC
injectors
A flow or pressure module can be connected to positions 1 – 8. Split vent modules or pressure modules
can only connect to positions 9 – 12.
The PPC modules for injectors 1 and 2 carrier gas connections are connected to sockets 1 (J10) and 2
(J11), respectively, on the PPC board.
The PPC modules for the split/splitless capillary (Cap) injector or PSS injector split vent connections that
correspond to injectors 1 and 2, or the pressure readout that corresponds to a packed column injector
(Pkd) or POC injector installed in socket 1 and 2 are connected to sockets 11 (J20) and 12 (J21),
respectively, on the PPC board.
The PPC modules for detectors 1 and 2 are connected to sockets 3 (J12) through 6 (J15) on the board.
Detector 1 uses sockets 3 and 4.
Содержание CLARUS 590 GC
Страница 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Страница 10: ...10 Clarus 590 GC User s Guide ...
Страница 11: ...1 Introduction ...
Страница 12: ...12 Clarus 590 GC User s Guide ...
Страница 21: ...2 Touch Screen Navigation ...
Страница 22: ...22 Clarus 590 GC User s Guide ...
Страница 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Страница 59: ...3 Using the Active Method ...
Страница 60: ...60 Clarus 590 GC User s Guide ...
Страница 109: ...4 Setting Up the Detectors ...
Страница 110: ...110 Clarus 590 GC User s Guide ...
Страница 197: ...5 Using the Method Editor ...
Страница 198: ...198 Clarus 590 GC User s Guide ...
Страница 207: ...6 Using the Tools Menu ...
Страница 208: ...208 Clarus 590 GC User s Guide ...
Страница 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Страница 225: ...Detectors 225 Output Configured ...
Страница 262: ...262 Clarus 590 GC User s Guide ...
Страница 263: ...7 Setting up a Typical Analysis ...
Страница 264: ...264 Clarus 590 GC User s Guide ...
Страница 279: ...Index ...
Страница 280: ...280 Clarus 590 GC User s Guide ...